Figure 1.
OBIC Image of a p-n Junction with
a
recombination center
The
spatial
resolution
of
single-photon backside OBIC imaging, however, is limited by the
compromise
between being
able to transmit the beam through the substrate and at the same time
allowing the beam to be absorbed by the semiconductor for
the generation of electron-hole pairs that are measurable as a current.
The range of
wavelengths that can meet both of these requirements does not give
single-photon OBIC enough resolution for effective use in analyzing IC's
with submicron IC features.
This
limitation of single photon OBIC in backside analysis may be overcome by
'two-photon absorption',
which involves two photons arriving at the same time on the sample to
coherently 'free' the electron. The photons used for this purpose must
have energies that are less than the band gap of the semiconductor, but
greater than half the band gap.
Two-photon absorption is a non-linear
form of absorption,
with the success of generating carriers
depending on the square of the light intensity. This
quadratic intensity
dependence makes two-photon carrier generation very inefficient outside
the area of focus. OBIC that operates on two-photon absorption is also
known as
'TOBIC', which stands for 'Two-photon OBIC.'
Applications
of OBIC
include: 1) detection of recombination or generation centers in a p-n
junction; 2) detection of inter-level shorts; 3) confirmation of the
'on' and 'off' states of a transistor; 4) detection of a latch-up
mechanism within a circuit; 5) location of weak points within a MOS
transistor; etc.