Atomic Force Microscopy (AFM)
Atomic Force Microscopy (AFM)
is a powerful form of scanning probe microscopy (SPM) that performs its imaging function by
measuring a local property of the surface being inspected, such as its
height, optical absorption, or magnetic properties.
Contact mode
imaging
employs a
soft cantilevered beam that has a sharp tip at its end, which is brought
in contact with the surface of the sample.
The force
between the tip and the sample causes the cantilever to deflect in
accordance with Hooke's Law, exhibiting a spring constant that typically
ranges between
The amount of deflection is measured by reflecting light from a laser diode off the back of the beam, and onto a pair or array of position-sensitive photodetectors. The differences between the reflected light received by the individual photodetectors indicate the amount of angular deflection of the cantilever at any given point on the sample. As a cheaper (but less sensitive) alternative to laser detection, the cantilever deflection may be measured using piezoresistive AFM probes that serve as a strain gauge system.
The ability
to monitor this deflection allows the AFM to create an image of the
sample non-destructively even if the tip is continuously in contact with
the sample. To prevent
the cantilever tip from damaging the surface of the sample, it is
maintained at a constant angular deflection so that the force applied by the tip on the surface is also kept
constant. This is achieved using a feedback mechanism that adjusts the
distance between the tip and the surface to keep the
applied force
constant.
The AFM's
feedback mechanism controls
piezoelectric elements
that hold the sample and move it in all
three axes relative to the tip. The z-movement is used for maintaining
the force at the right level, while the x- and y- movements are used for the
raster-scanning the tip over the sample. The resulting map
Non-contact
imaging
The advantages of AFM over electron microscopy include the following: 1) it generates true, 3-dimensional surface images; 2) it does not require special sample treatments that can result in the sample's destruction or alteration; and 3) it does not require a vacuum environment in order to operate (it can operate in both air and liquid). On the other hand, its disadvantages include the following: 1) the image size that it provides is much smaller than what electron microscopes can create; and 2) it is slow in scanning an image, unlike an electron microscope which does it in almost real-time.
In the semiconductor industry, AFM is primarily used for the imaging of VLSI cross-sections. Materials that can be imaged by AFM include metals, polymers, photoresists, etc. See also: STM.